Frag' FlorenceEvidenz. Klar. Anwendbar.
Uhr 7/8Sources Journal Tree
Easy Demo

Lokaler Crossref-Datenbestand · journal-article

5-year outcomes of iStent micro-bypass vs. Kahook dual blade goniotomy in open angle glaucoma

Ejaz Ansari

Graefe's Archive for Clinical and Experimental Ophthalmology · 2026

Vollständiger Abstract

Worum geht es in dieser Arbeit?

Abstract Purpose Long-term comparison of efficacy and safety of i-Stent versus KDB goniotomy combined with phacoemulsification through to 5 years. Methods A retrospective chart review of 88 patients that received iStent or KDB in combination with phacoemulsification at a single centre between January 2014 and January 2019. Outcome measures were proportion of eyes achieving ≥ 20% reduction of IOP, drop dependence through to 5 years and change in visual field mean deviation (MD) at year 5 compared to baseline. Early to moderate stage of glaucoma. Results No significant difference between groups for IOP reduction ( p = 0.3) or visual field MD ( p = 0.85) at all time points through to 5 years. Significantly lower drop dependence favouring KDB at all time points ( p < 0.001 year 1; p = 0.02 year 5). Lower rate of secondary procedures following KDB (7.7% vs. 14% p = 0.48). Adverse events following iStent and KDB: hyphaema (5.5% v 10% p = 0.70), corneal oedema (8% v 2% p = 0.30), rebound anterior uveitis (5.5% v 2% p = 0.57). Complications resolved within 2 weeks with no serious incidents (loss of VA by ≥ 2 Snellen lines/ 0.3 logMAR). Conclusion Each procedure provided comparable long-term IOP reduction and visual field stability over five years. KDB goniotomy resulted in significantly less drop dependence throughout with fewer associated secondary procedures. Adverse events were infrequent and mild, with no serious complications observed.

Bibliografischer Nachweis

Publikationsdaten

Autor:innen
Ejaz Ansari
Quelle
Graefe's Archive for Clinical and Experimental Ophthalmology
Publikation
2026-01-01
Band / Ausgabe
Nicht angegeben
Seiten
Nicht angegeben
ISSN / ISBN
0721-832X, 1435-702X
Zitationen
0 laut Crossref
Referenzen
0 hinterlegt

Zitieren

Zitierfähiger Nachweis

Ejaz Ansari (2026). 5-year outcomes of iStent micro-bypass vs. Kahook dual blade goniotomy in open angle glaucoma. Graefe's Archive for Clinical and Experimental Ophthalmology. https://doi.org/10.1007/s00417-026-07372-4
RIS BibTeX CSL-JSON

Kontext

Themen, Förderung und Nutzung

Lizenzhinweise: Lizenz 1