Vollständiger Abstract
Worum geht es in dieser Arbeit?
Background Healthcare workers in critical care settings are at increased risk of acquiring respiratory infections, particularly during aerosol-generating procedures (AGPs). Portable negative-pressure isolation devices have been proposed to fix airborne infection in isolation rooms. Objective To evaluate whether an isolated negative-pressure device reduces bacterial and fungal aerosol dispersion under controlled laboratory conditions. Methods This laboratory-based controlled experimental observational study assessed aerosol containment across three techniques: (1) no cover, (2) plastic sheet canopy barrier, and (3) a portable negative-pressure isolation system. Standardized 5 mL suspensions of Staphylococcus epidermidis and Aspergillus flavus were nebulized on three separate days for each technique. Nutrient agar plates were placed and colony-forming units (CFU) were manually counted. The primary outcome was the mean CFU per technique per day. One-way ANOVA with post-hoc Welch’s t -tests (Bonferroni-corrected) was used for statistical comparison. Results A total of 36 bacterial and fungal plates were analysed. For bacterial aerosols, mean CFU was highest with no cover (535.6 ± 5.38), lower with plastic sheet canopy barrier (429.5 ± 5.41), and lowest with the negative-pressure system (217.5 ± 17.5). Differences were statistically significant (ANOVA p = 9.8 × 10 −8 ; η 2 = 0.995). Fungal CFU demonstrated a similar pattern with significantly lower CFU for Technique 3 compared with both other techniques ( p < .001). Conclusion The negative-pressure isolation device significantly reduced airborne dispersion of bacterial and fungal aerosols compared with plastic sheet canopy barrier methods. These findings support the use of portable negative-pressure systems to enhance aerosol containment during AGPs. Further clinical validation is recommended.
Bibliografischer Nachweis
Publikationsdaten
- Autor:innen
- Sunit Kumar Gupta, Nidhima Aggarwal, Nisha M. P, Vamshidhar Chamala, Samarjit Dey, Eeshwar M. V, Sumit Bansal, Mukesh Tripathi, Jitendra Chawla
- Quelle
- Journal of Infection Prevention
- Publikation
- 2026-01-01
- Band / Ausgabe
- Nicht angegeben
- Seiten
- Nicht angegeben
- ISSN / ISBN
- 1757-1774, 1757-1782
- Zitationen
- 0 laut Crossref
- Referenzen
- 0 hinterlegt
Zitieren
Zitierfähiger Nachweis
Sunit Kumar Gupta, Nidhima Aggarwal, Nisha M. P, Vamshidhar Chamala, Samarjit Dey, Eeshwar M. V, Sumit Bansal, Mukesh Tripathi, Jitendra Chawla (2026). Can isolated negative-pressure devices reduce transmission of respiratory infection in critical care units: A laboratory-based controlled experimental observational study. Journal of Infection Prevention. https://doi.org/10.1177/17571774261477869
Kontext
Themen, Förderung und Nutzung
Lizenzhinweise: Lizenz 1