A commissioning workflow for quantitative ¹⁷⁷Lu SPECT/CT across different Siemens scanner–reconstruction configurations

Leonardo Lampertico, Gaia Muti, Claudia Carbonini, Chiara Romanò, Antonio Scarale, Claudio Rossetti, Paola Enrica Colombo

Advances in Radiotherapy & Nuclear Medicine · 2026 · Band 0 · Ausgabe 0 · S. 026260029

Vollständiger Abstract

Worum geht es in dieser Arbeit?

The accuracy of quantitative lutetium-177 single-photon emission computed tomography/computed tomography (¹⁷⁷Lu SPECT/CT) depends on the imaging system, acquisition protocol, reconstruction algorithm, and calibration strategy. This study aimed to develop and evaluate a practical commissioning workflow to optimize ¹⁷⁷Lu imaging across different commercial scanner–reconstruction configurations. Three Siemens workflows were evaluated: FLASH3D and xSPECT Quant on the Symbia Intevo Bold, and FLASH3D+ on the Symbia Pro.specta. A uniform phantom was used to assess image noise and derive calibration factors for the ordered-subset expectation maximization (OSEM)-based reconstructions. Recovery coefficients (RCs) were evaluated via an International Electrotechnical Commission/National Electrical Manufacturers Association phantom at various sphere-to-background ratios and modeled using a three-parameter logistic function for partial volume correction. Reconstruction parameters were optimized by jointly evaluating RC convergence and image noise, followed by quantitative validation using an anthropomorphic phantom. OSEM-based workflows showed a predictable dependence on equivalent iterations, yielding lower image noise. Conversely, xSPECT Quant exhibited complex parameter dependence and higher noise, but provided superior activity recovery—with RCs closer to unity—and the lowest quantification errors during validation. A reconstruction setting of two subsets and 30 iterations was selected for all workflows. For FLASH3D and FLASH3D+, this configuration provided RCs closest to unity while maintaining the coefficient of variation below the predefined 15% threshold. The same setting was adopted for xSPECT Quant to ensure methodological consistency despite its intrinsically higher image noise. This commissioning workflow provides a practical framework for local ¹⁷⁷Lu SPECT/CT optimization. Among the configurations, xSPECT Quant yielded higher recovery coefficients and lower quantification errors, albeit with increased noise. These findings highlight the need for locally optimized protocols for patient-specific dosimetry and do not establish the general superiority of any one reconstruction algorithm.

Bibliografischer Nachweis

Publikationsdaten

Autor:innen
Leonardo Lampertico, Gaia Muti, Claudia Carbonini, Chiara Romanò, Antonio Scarale, Claudio Rossetti, Paola Enrica Colombo
Quelle
Advances in Radiotherapy & Nuclear Medicine
Publikation
2026-08-20
Band / Ausgabe
Nicht angegeben
Seiten
026260029
ISSN / ISBN
3060-8554, 2972-4392
Zitationen
0 laut Crossref
Referenzen
0 hinterlegt

Zitieren

Zitierfähiger Nachweis

Leonardo Lampertico, Gaia Muti, Claudia Carbonini, Chiara Romanò, Antonio Scarale, Claudio Rossetti, Paola Enrica Colombo (2026). A commissioning workflow for quantitative ¹⁷⁷Lu SPECT/CT across different Siemens scanner–reconstruction configurations. Advances in Radiotherapy & Nuclear Medicine, 0 (0), 026260029. https://doi.org/10.36922/arnm026260029
RIS BibTeX CSL-JSON

Kontext

Themen, Förderung und Nutzung

Lizenzhinweise: Lizenz 1